Microfluidizer Á¾ÇÕÄ«´Ù·Î±×


Phone
  ´ëÇ¥ÀüÈ­ 1577-7956  
  º»     »ç 031-457-9187  
  ´ë     Àü 042-932-1265

Web Site Visitors
  Today 157    
  Yesterday 767    
  Since 2006 1,763,437    


 
¸®½ºÆ®µ¹¾Æ°¡±â

Industry : Food
Product : Oil and water emulsion
Test Objective : Reduce particle size to 0.5 um
Chambers : - Interaction Chamber (upstream) H230Z (400 um)
- Auxiliary Processing Module (downstream) H210Z (200 um)
Processor : M-110Y     (Microfluidizer High Shear Processor Àåºñº¸±â)
Number of passes : 1
Shear per pass : 830,000 sec-1
Process pressure : 3,000 psi
Particle size : - Premix : 5.97 um mean particle size [D90 = 12.5 um]
- 1 Pass : 0.61 um mean particle size [D90 = 0.98 um]
Summary of results : Customer's previous homogenization method reduced the droplet size to 5.0 um. The M-110Y Microfluidizer processor reduced the mean droplet size to 0.607 um after 1 pass at 3,000 psi
(pdf È­ÀÏ ´Ù¿î·Îµå)












 
(ÁÖ)¿µÁøÄÚÆÛ·¹ÀÌ¼Ç   |   ´ëÇ¥ : ÀÌÇü¿­   |   »ç¾÷ÀÚµî·Ï¹øÈ£ : 138-81-34745
±º Æ÷ º» »ç   |   (15853) °æ±âµµ ±ºÆ÷½Ã ÇѼ¼·Î 48 ³²¿øºôµù 2Ãþ   |   ÀüÈ­ : 1577-7956 (1)   |   Æѽº : 031-457-9188   |   À̸ÞÀÏ : info@yjcorp.co.kr
´ë Àü Áö »ç   |   (34036) ´ëÀü½Ã À¯¼º±¸ Å×Å©³ë10·Î 29 ÆÄÅ©ÆÓ¸®½º 2Ãþ   |   ÀüÈ­ : 1577-7956 (2)   |   Æѽº : 042-932-1266   |   À̸ÞÀÏ : info@yjcorp.co.kr