Industry : Food
Product : Oil and water emulsion
Test Objective : Reduce particle size to 0.5 um
Chambers : - Interaction Chamber (upstream) H230Z (400 um)
- Auxiliary Processing Module (downstream) H210Z (200 um)
Processor : M-110Y     (Microfluidizer High Shear Processor Àåºñº¸±â)
Number of passes : 1
Shear per pass : 830,000 sec-1
Process pressure : 3,000 psi
Particle size : - Premix : 5.97 um mean particle size [D90 = 12.5 um]
- 1 Pass : 0.61 um mean particle size [D90 = 0.98 um]
Summary of results : Customer's previous homogenization method reduced the droplet size to 5.0 um. The M-110Y Microfluidizer processor reduced the mean droplet size to 0.607 um after 1 pass at 3,000 psi