| Specifications |
OmniaTap |
OmniaTap UV-TOC |
OmniaTap UV-TOC/UF |
|
| Pure water values type II |
|
|
| Pure water performance at 15 ¡ÆC [l/h] |
12 or 20 |
12 or 20 |
12 or 20 |
|
| Conductivity at 25 ¡ÆC [¥ìS/cm] |
0.067 up to 0.1 |
0.067 up to 0.1 |
0.067 up to 0.1 |
|
| Resistance at 25 ¡ÆC [M¥Ø x cm]
| 15 up to 10 |
15 up to 10 |
15 up to 10 |
|
|   |
|
| Ultrapure water values type I |
|
|
| Conductivity at 25 ¡ÆC [¥ìS/cm] |
0.055 |
0.055 |
0.055 |
|
| Resistance at 25 ¡ÆC [M¥Ø x cm]
| 18.2 |
18.2 |
18.2 |
|
| TOC-value* [ppb] |
< 10 |
< 5 |
< 5 |
|
| TOC monitor |
no |
yes |
yes |
|
| Dispensing performance [l/min.] |
up to 2 |
up to 2 |
up to 2 |
|
| Individually adjustable dispensing volume [liters] |
0.05 up to 25 |
0.05 up to 25 |
0.05 up to 25 |
|
| Particles** > 0.2 ¥ìm [1/ml] |
< 1 |
< 1 |
< 1 |
|
| Bacteria** [CFU/ml] |
< 0.01 |
< 0.01 |
< 0.01 |
|
| Pyrogens (Endotoxins)*** [EU/ml] |
- |
- |
< 0.001 |
|
| RNase*** [pg/ml] |
- |
- |
< 1 |
|
| DNase*** [pg/ml] |
- |
- |
< 5 |
|
| Proteases*** [¥ìg/ml] |
- |
- |
< 0.15 |
|
* The values given are typical and may vary depending on the quality of the feed water
** With sterile filter capsule 0.2 ¥ìm or bio filter capsule
*** With ultrafilter/bio filter capsule
|
|
| Feedwater requirements |
|
|
| Tap water according to DIN 2000 |
|
| Feedwater pressure [bar] |
1 up to 6 |
1 up to 6 |
1 up to 6 |
|
| Conductivity at 25 ¡ÆC [¥ìS/cm] |
< 2000* |
< 2000* |
< 2000* |
|
| Colloid index SDI |
< 5** |
< 5** |
< 5** |
|
| Dissolved CO©ü [ppm] |
< 30 |
< 30 |
< 30 |
|
| Free chlorine [ppm] |
< 0.1 |
< 0.1 |
< 0.1 |
|
| TOC [ppb] |
< 2 |
< 2 |
< 2 |
|
| Hardness [as CaCO3] [ppm] |
< 300 |
< 300 |
< 300 |
|
| Iron/manganese [mg/l] |
< 0.05 |
< 0.05 |
< 0.05 |
|
| Silica [ppm] |
< 30 |
< 30 |
< 30 |
|
| pH range |
4 up to 10 |
4 up to 10 |
4 up to 10 |
|
* Feed water with high conductivity can reduce the service life of the cartridges and increase the conductivity of type III water.
  If the conductivity is between 800 and 2000 ¥ìS/cm, we recommend using a water softener
** With an SDI/FI between 3 and 5, pre-treatment must be used |
|
| Technical data |
|
|
| Feedwater connection |
R3/4 inch |
R3/4 inch |
R3/4 inch |
|
| Electrical connection [Volt/Hz] |
90 - 240/50 - 60 |
90 - 240/50 - 60 |
90 - 240/50 - 60 |
|
| Connected load [W] |
0.1 |
0.1 |
0.1 |
|
| Ambient temperature [¡ÆC] |
4 up to 40 (Recommendation: 10 up to 25) |
|
| Dimensions without tank* [WxHxD mm] |
390 x 720 x 525 |
390 x 720 x 525 |
390 x 720 x 525 |
|
| Dimensions with 10 L docking tank* [WxHxD mm] |
390 x 720 x 615 |
390 x 720 x 615 |
390 x 720 x 615 |
|
| Weight without 10-L docking tank [kg] |
17 |
18 |
18 |
|
| Weight with 10-L docking tank [kg] |
20 |
21 |
21 |
|
| * With OptiFill Dispenser |