| Specifications |
OmniaLabRO 20     |
OmniaLabRO 40     |
OmniaLabRO 60     |
|
| Reverse Osmosis Water Type III |
|
| Pure water performance at 15 ¡ÆC [l/h] |
20 |
40 |
60 |
|
RO membrane retention rate in %
(ions. germs and bacteria) |
> 98 |
> 98 |
> 98 |
|
| |
|
| Feedwater requirements |
|
|
| Tap water according to DIN 2000 |
|
| Feedwater pressure [bar] |
1 up to 6 |
1 up to 6 |
1 up to 6 |
|
| Conductivity at 25 ¡ÆC [¥ìS/cm] |
< 2000* |
< 2000* |
< 2000* |
|
| Colloid index SDI |
< 5** |
< 5** |
< 5** |
|
| Dissolved CO©ü [ppm] |
< 30 |
< 30 |
< 30 |
|
| Free chlorine [ppm] |
< 0.1 |
< 0.1 |
< 0.1 |
|
| TOC-value [ppb] |
< 2 |
< 2 |
< 2 |
|
| Hardness [as CaCO3] [ppm] |
< 300 |
< 300 |
< 300 |
|
| Iron/manganese [mg/l] |
< 0.05 |
< 0.05 |
< 0.05 |
|
| pH range |
4 up to 10 |
4 up to 10 |
4 up to 10 |
|
* Feed water with high conductivity can reduce the service life of the cartridges and increase the conductivity of type III water.
  If the conductivity is between 800 and 2000 ¥ìS/cm, we recommend using a water softener
** With an SDI/FI between 3 and 5, pre-treatment must be used |
|
| Technical data |
|
|
| Feedwater connection |
R3/4 inch |
R3/4 inch |
R3/4 inch |
|
| Electrical connection [Volt/Hz] |
90 - 240/50 - 60 |
90 - 240/50 - 60 |
90 - 240/50 - 60 |
|
| Connected load [W] |
0.1 |
0.1 |
0.1 |
|
| Ambient temperature [¡ÆC] |
4 up to 40 (Recommendation: 10 up to 25) |
|
| Dimensions Tower* [WxHxD mm] |
511 x 1520 x 575 |
511 x 1520 x 575 |
511 x 1520 x 575 |
|
| Dimensions Base cabinet tank [WxHxD mm] |
511 x 800 x 575 |
511 x 800 x 575 |
511 x 800 x 575 |
|
| Weight [kg] |
38 |
36 |
39 |
|
| * With OptiFill Dispenser |